With the Dose Calibrator ISOMED 2000, MED Nuklear-Medizintechnik has created an instrument combining highest performance and small footprint. The measuring system has been designed for complex applications in the clinical/ university field und in PET and research centers.
Measuring range: | |
e.g. Tc-99m | 40 kBq to 50/200* GBq (* 7% additional error) |
F -18 | 60 kBq to 70/300* GBq |
Measuring range setting | automatically, alternative:fixed measuring range |
adjustable e.g. for PET filling | |
Energy range for Gamma sources | 25 keV to 3 MeV |
Measurement time | |
with meas. range change-over | 2s – 15s |
without meas. range change-over | 1s – 3s |
Basic error | < 5% |
Linearity error | < 2% |
Result display | 4-digits including display of unit, nuclide and chemical compound |
Stored isotope table | C-11, N-13, O-15, F-18, P-32, Cr-51, Mn-54, Co-57, Co-58, Fe-59, Co-60, Ga-67, Se-75, Sr-89, |
Y-90, Mo-99, Tc-99m, In-111, In-113m, I-123, I-125, I-131, Xe-133, Cs-137, Ba-140, Sm-153, | |
Er-169, Yb-169, Re-186, Re-188, Hg-197, TI-201, Ra-224 | |
Containers: | |
Injector syringe | 1, 2, 3, 5, 10, 20 ml |
Bottles | 5, 10 (P6), 15, 20 ml |
Ampoules | 5 ml |
Capsules | for Co-57, Co-58, I-131 |
Contents (sample quantity) | 0.1 ml – 99.9 ml |
Measurement chamber | chamber ø 125/184 mm, well ø 47 mm |
Dimensions | total height 387 mm, well depth 205 mm |
Shielding | 4 mm Pb basic shielding, additional shielding 16 or 50 mm |
PC-System | industry-standard PC-system (PC 104) in flat housing ( 250 x 330 x 55 mm) |
Monitor | LC-display 12,1” TFT |
Keyboard | touch-screen, external keyboard or mouse |
Weight | approx. 10 kg (without additional shielding) |