MULTICAM 1000 / MULTICAM 2000

MULTICAM 1000 / MULTICAM 2000
• Number of detector heads 2 for MultiCam 2000
  1 for MultiCam 1000

Detector

,
• UFOV dimensions UFOV: 381 mm x 508 mm
• Crystal thickness 9.5 mm
• Exit window Glass: 15.9 mm
• Crystal size 470 mm x 593 mm
• Number of PMT’s 55 tubes per detector:
49 76.2 mm (3") round tubes
6 50.8 mm (2") round tubes
• PMT shielding Yes, high permeability Mu-metal
magnetic shields on every PMT
Maximum magnetic operating field: .5 gauss
• FOV to detector edge 9.0 cm
• Detector shielding 4 sides shielded with 19.0 mm lead minimum,
base shielding of 31.8 mm lead
• PMT Mounting Directly on the crystal output window- no additional light pipe used
• Energy range 30 - 800 KeV
• Type of detector motion Radial and tilt motion
• Detector movement Individually or simultaneously, either direction
(maximum of two simultaneous motions)
• Iso-center (92 cm) from floor
• Distance of crystal face from floor Face up = 58 cm to 80 cm
Face down = 105 cm to 127cm
• Total radial travel 21 .0 cm
• Total range 13.0 cm - 34.00 cm
Measured from crystal surface to center of rotation.
9.83 cm - 30.8 cm
Measured from LEHR collimator surface to COR
• Radial speed Acquisition
Maximum: 60 cm/min
Minimum: 3 cm/min
Manual (hand-held controller)
Maximum: 60 cm/min
Minimum: 30 cm/min
• Radial position resolution Encoded to 0.007 mm
• Radial position accuracy +/- 0,8 mm
• Detector tilt range - 1° to + 90°
• Tilt speed Manual (hand-held controller)
Maximum: 0.50 rpm
Minimum: 0.125 rpm
• Tilt Resolution Encoded to 0.01 °
• Tilt accuracy ±0.1 ° @ 0° ± 0.2° over entire r ange (-1 ° - +91 ° )
• Detector locations 180° opposed geometry
• Detector registration Proprietary Image Registration Correction (IRC) aligns all
heads to the axis of rotation and to each other through
calibration using three point sources

Collimator

• Fabrication process Foil
(Cast - S:O.I. Required
Not available in all types of collimators)
• Automatic collimator identification Yes
• Collimator exchange server Yes, vertical server
• Server capacity 113 kg maximum
• Collimator storage device Vertical storage
• Collision sensing device Yes
• Storage capacity Ouantity - 4 (standard)
113 kg maximum per slot
Optional second storer can be placed adjacent to first storer.
(see page 2-10 for types of collimators/specifications)

Gantry

• Type Ring
• Diameter of aperture
(scan diameter without collimator)
Maximum: 68 cm
Minimum: 26 cm
• Diameter of aperture
(scan diameter with
LEHR collimator)
Maximum: 64.8 cm
Minimum: 22.8 cm
• Rotation 400° total rotation
• Angle resolution Encoded to 0.01 °
• Rotation accuracy ±0.1 °
• Rotation modes Continuous and step-and-shoot
• Speed of rotation Acquisition
Maximum: 1 .0 rpm
Minimum: 0.025 rpm
Manual (hand-held controller)
Fast speed: up to 1 .0 rpm
Slow speed: as slow as 0.25 rpm
• Rotation step time 4.0 sec/3 degrees for circular orbit
• Rotation increments Selectable in multiples of 1 degree
• Whole body scan Yes
• Gantry whole body scan travel 0 cm - 176 cm (optional 0cm – 198 cm)
• Total whole body imaging length 198 cm (optional 220 cm)
• Whole body acquisition speed Maximum: 300 cm/min
Minimum: 3 cm/min
• Operator programmable
WB contouring
Automatic with light curtain
• Speed - manual .
(hand-held controller)
Maximum 300 cm/min
Minimum: 120 cm/min
• Digital display of gantry
and detector movements
Yes , on color TFT-Display
• What type of information is
displayed to the operator
Gantry:   rotation angle, linear translation
Detector:   tilt angle, radial position
Table:   Distance from center of table top surface to iso-center
Persistence imaging in real time
Error information
Help text to prompt operator
Information on collimator status
System status information. (emergency stop, collision, etc..)
Collimator type
• Gantry control accessible
to the operator
Yes, through operator hand-held controller
• Preprogrammed functions Yes
Whole body study setup
Brain SPECT study setup
Torso (heart) SPECT study setup
Upright study setup
Patient load/unload
Collimator load

Patient Table

during acquisition for heart centering
• Type Non-cantilevered type (fixed supports on both ends)
• Material Carbon-fiber
• Dimensions of the pallet 244 cm long, 41 cm wide (optional 259 cm long, 50 cm wide)
• Useable dimension 210 cm long, 41 cm wide (optional 225 cm long, 50 cm wide)
• Maximum patient weight load 159 kg (optional 200Kg)
• Attenuation information 9° attenuation at 140 Ke V
• Vertical movement 20.0 cm (64 cm - 84 cm measured from the floor)
• Minimum distance to the floor 64 cm
• Vertical speed Manual (hand-held controller)
Fast speed: 60 cm/min
Slow speed: 30 cm/min
(with maximum load)
• Horizontal movement None (gantry moves instead)
• Lateral movement 5.08 cm manually
• Lateral motion accuracy +/- 1 mm
• Control to the patient table
Accessible to the operator
Through hand-held control
(except lateral movement which is manual)
• Head rest Optional carbon fiber head holder with foam inserts for
head restraint also available
• Thickness of table 1.9 cm (314") (in center of table)
• Table weight 8.5 kg (optional 10,5 Kg)

Pedriatic Patient Table

• Type Non-cantilevered type (fixed supports on both ends)
• Material Carbon-fiber
• Dimensions of the pallet This pallet has a specially designed pediatric area which is
contoured and narrower in shape than the standard table. It
allows optimum SPECT imaging as detectors stay closer to
smaller pediatric patients
• Usable dimension 210 cm
• Maximum patient weight load 68 kg
• Attenuation information <9%
• Vertical movement 20.0 cm (64 cm - 84 cm measured from the floor)
• Minimum Distance to the floor 64 cm
• Vertical Speed Manual (hand held controller)
Fast speed: 60 cm/min
Slow speed: 30 cm/min
(with maximum load)
• Horizontal movement
during acquisition
None (gantry moves instead)
• Lateral motion accuracy ± 1 mm
• Control to the patient table
accessible to the operator
Yes, through hand-held controller
(except lateral movement which is manual)
• Head rest Yes
• Thickness of table 1.9 cm ( in center of table)
• Table weight Approx. 6.8 kg

Safety Features

• Emergency stop switches Three on gantry Yes, membrane sensing device standard
with each collimator (except pinhole collimator)
• Collision sensing. detection Yes
• Head retract function for detectors Yes
• Collimator latch sensor Yes
• Sensor for proper collimator mounting Yes
• Regulatory compliance tested to IEC 60601-1 specifications
• Quality Compliance ISO 9000 compliant

Detector Specificafions
NEMA Performance Specifications*

Performance Characteristics** Worst Case Values
3/8" (9.5mm)
Intrinsic Energy Resolution (IER) 10.0%
Intrinsic Spatial Resolution (ISR) CFOV FWHM 3.8 mm
FWTM 7.2 mm
UFOV FWHM 3.9 mm
FWTM 7.4 mm
Intrinsic Count Rate Perf. – full corr.
with 20% energy window;
Max. Count Rate CPS 320K
Observed 20%
Count Loss Rate
CPS 150K
Maximum Count Throughput ***  650 K
Intrinsic Flood Field Uniformity (IFFU) CFOV DIFF 1.9 %
INTEGRAL 2.0 %
Intrinsic Spatial Linearity (ISL) UFOV DIFF 2.4 %
INTEGRAL 2.5 %
CFOV DIFF 0,2 mm
INTEGRAL 0,5 mm
System Spatial Resolution
CFOV at 10cm
LEHR Collimator
No Scatter FWHM 8.2mm
FWTM 14.5 mm
With Scatter FWHM 8.7 mm
FWTM 21.2 mm
System Sensitivity TC-99m
With LEHR
Collimator
Cts/min/Ci 245
Point Source Sensitivity UFOV  +/- 5%
Intrinsic Spatial Resolution
(ISR) at 75K CPS
UFOV FWHM 3.9
FWTM 7.5
Intrinsic Flood Field Uniformity CFOV DIFF 2,5%
INTEGRAL 3.0%
UFOV DIFF 2.5%
INTEGRAL 3.0%
* Performance specifications for 15.0" x 20.0" rectangular detector with 3/8" Nal crystal NEMA NU1-1986
   Performance Measurement of Scintillation Cameras
**All performance characteristics measured in Normal mode except where noted

Collimator Specifications

Description Field of
View at
100mm
(mm)
MAX.
Useful
Energy a
(Kev)
Hole
Size b
Hole
Length
(mm)
Septa
Thickness
(mm)
System
Resol. at
0mm
(mm)
System
FWHMc
at
100mm
(mm)
System
Sensitivity d
(cpm/uCi)
Dynamic (DY) 381 x 508 150 2.31 25.4 0.30 4.8 13.4 740
General All Purpose
(GAP)
By SOI only
381 x 508 150 1.57 25.4 0.24 4.3 9,5 337
High Resolution (HR) 381 x 508 160 1.40 27.0 0.18 4.2 8.3 245
Ultra High Resolution
(UHR)
381 x 508 160 1.40 34,9 0,15 4,2 7,1 155
High Resolution
Fanbeam(HR-Fan)
381 x 380
Focus=5Ocm
160 1.40 27,0 0,18 4,2 8,3 300
High Resolution
Fanbeam (Cardio Fan)
381 x 412
Focus=65cm
160 1.40 27,0 0,18 4,2 8,3 280
Ultra High Resolution
Fanbeam (UHR-Fan)
381 x 380
Focus=5Ocm
160 1,40 34,9 0,15 4,2 7,1 185
Medium Energy (ME) 381x508 300 3,40 58,4 0,86 5,3 11,2 220
High Energy (HE) 381x508 400 3,81 58,4 1,73 5,8 12,7 220
Pinhole (3 Apertures) 145 dia. 400   - -      
    3,0     5,1 6,4  
    5,0     6,8 9,3  
    7,0     8,6 12,5  
a) Based on approximately 5% Septal Penetration.

Physical Specifications

Components Weight
(kg)
Gantry/Table Assembly (MultiCam 2000) 1650
Gantry/Table Assembly (MultiCam 1000) 1350
Collimator Storer (without collimators) 69
Collimator Server 59
UniCam Computer 12
Display Monitor 15
Collimators 113 max.

Power Requirements

Components Power
Gantry Assembly include Computer 230 V, 16 Amp, Single Phase (60Hz/50Hz)

Source Configuration

The power source supplying MULTICAM system shall be single phase, 2-wire plus dedicated/isolated ground and neutral (4-wire). The source should be sized for 10 KVA load.

Nominal AC Volts 230
Line Phase Single
Voltage Regulation 5%
Full Load kVA 1,5
Line Amps 6
Voltage Variation +5% -5%
Distribution Transformer 1,5 kVA

Quality of Power

Power quality specifications are based on measurements taken at the Nuclear Room power distribution panel.

Voltage Fluctuation (Sags/Surges)

Sags and surges in the line shall not cause the line-to-line or line voltage to deviate from either its nominal or average value by more than 7%, whichever is less.

Brownouts

If there are consistent voltage losses of greater than 8 cycles (120 msecs) a battery UPS is recommended.

Environmental Specification

Component Heat Output Component
(KiloWatt)
Gantry Assemblies
(includeComputer)
0,7 max.
MULTICAM 1000 / MULTICAM 2000 space requirements