Exradin A12 Ion Chamber

92700 A12

Exradin A12 Ion Chamber

UNCOMPROMISING QUALITY

For absolute dosimetry calibrations in water, air, or other phantom material

DESIGNED FOR ABSOLUTE DOSIMETRY CALIBRATION

Exradin Farmer-type Chambers are specifically designed for absolute dosimetry calibrations in water, air, or other phantom material. The Model A12 is completely characterized in TG 51 and TRS 398.

FAST, PRECISE MEASUREMENTS

Its waterproof construction and two piece removable stem makes it ideal for use in water phantoms. The chamber vents through a flexible tube that surrounds the triaxial cable, ensuring the collecting volume is in pressure equilibrium with the surroundings. The design assures there are no stem or voltage soakage effects, providing precise and reliable measurements.

DURABLE CONSTRUCTION, BUILT TO LAST

Farmer-type Chambers are constructed of rugged C552 Shonka air-equivalent plastic, providing excellent conductivity and years of reliable use. Rigorous one meter drop test proves ruggedness and reliability.


Specifications

REF Number 92700
Collecting Volume 0.65 cm3
Nominal Calibration Factor 5 R/nC
Centroid of Collecting Volume 12.9 mm from tip of chamber
Collector Diameter 1.0 mm
Outside Diameter of Shell Collecting Volume 7.1 mm
Wall Thickness 0.5 mm
Shell, Collector, and Guard Material Shonka air-equivalent plastic C552
Included 60Co Buildup Cap Wall thickness of 2.8 mm; contructed of C552
Venting Vented to the ambient via the flexible vent tube surrounding the triaxial cable
Stem 1.3 cm OD black phenolic, two-piece with 5.1 and 12.7 cm segments
Nominal Collection Efficiency 1.000
Maximum Polarizing Potential 1000 V
Nominal Inherent Leakage Currents 10-15 A
Low-Noise Triaxial Cable 50 ohms, 29 pF/ft, 1.5 m long
Signal Connector Triaxial BNC plug (2-Lug, male pin); others available upon request
High Voltage Connector Integral with triaxial connector (shell of chamber is common with connector body)
Waterproof Yes
Product Standards CE0413, Designed to meet IEC 60601-1, IEC 60731


Model 12 Schematic


Specifications subject to change without notice.