For assessing pinpoint radiation fields for IMRT, stereotactic, orthovoltage x-rays, stereotactic, and superficial skin therapy
The Model 14 Microchamber is designed specifically for assessing radiation fields encountered in IMRT and stereotactic radiosurgery. Outwardly, the Microchamber is identical to the Model 1 Exradin Miniature Shonka Thimble chamber; both use the same base and shell. However, the Microchamber is capable of measuring extremely small field sizes of 4 mm by 6 mm allowing for exceptional spatial resolution and exact pinpoint beam profile characterization.
Its waterproof construction and two piece removable stem makes it ideal for use in water phantoms. The chamber vents through a flexible tube that surrounds the triaxial cable, ensuring the collecting volume is in pressure equilibrium with the surroundings. The design assures there are no stem or voltage soakage effects, providing precise and reliable measurements.
The Model 14 Microchamber is offered in two versions of rugged, Shonka plastic, providing excellent conductivity and years of reliable use:
REF Number | 92711 - Model A14 92712 - Model T14 |
Collecting Volume | 0.016 cm3 |
Nominal Calibration Factor | 365 R/nC |
Centroid of Collecting Volume | Approximately 2.0 mm from tip of chamber |
Collector Diameter | 0.33 mm |
Outside Diameter of Shell Collecting Volume | 6.0 mm |
Wall Thickness | 1.0 mm |
Shell, Collector, and Guard Material | A14 - Shonka air-equivalent plastic C552, wire, conductive plastic T14 - Shonka tissue-equivalent plastic A150 |
Venting | Vented to the ambient via the flexible vent tube surrounding the triaxial cable |
Stem | 1.3 cm OD black phenolic, two-piece with 5.1 and 12.7 cm segments |
Nominal Collection Efficiency | 1.000 |
Maximum Polarizing Potential | 1000 V |
Nominal Inherent Leakage Currents | 10-15 A |
Low-Noise Triaxial Cable | 50 ohms, 29 pF/ft, 1.5 m long |
Signal Connector | Triaxial BNC plug (2-Lug, male pin); others available upon request |
High Voltage Connector | Integral with triaxial connector (shell of chamber is common with connector body) |
Waterproof | Yes |
Product Standards | CE0413, Designed to meet IEC 60601-1, IEC 60731 |