For assessing pinpoint radiation fields for IMRT, stereotactic, orthovoltage x-rays, stereotactic, and superficial skin therapy
The Model A14SL Microchamber is designed specifically for assessing radiation fields encountered in IMRT and stereotactic radiosurgery. The Model A14SL thimble chamber is the Slim-Line version of Model A14. It has the exact same internal dimensions and collecting volume as the Model A14, yet the entire chamber has the uniform diameter of 6.4 mm (0.250 in), which is ideal when fitting it into a small phantom. The Microchamber is capable of measuring extremely small field sizes of 4 mm by 6 mm, allowing for exceptional spatial resolution and exact pinpoint beam profile characterization.
The chamber vents through a flexible tube that surrounds the triaxial cable, ensuring the collecting volume is in pressure equilibrium with the surroundings. The design assures there are no stem or voltage soakage effects, providing precise and reliable measurements.
The Model A14SL is constructed of rugged C552 Shonka air-equivalent plastic, providing excellent conductivity and years of reliable use.
REF Number | 92723 |
Collecting Volume | 0.016 cm3 |
Nominal Calibration Factor | 365 R/nC |
Centroid of Collecting Volume | Approximately 2.1 mm from tip of chamber |
Collector Diameter | 0.33 mm |
Outside Diameter of Shell Collecting Volume | 6.35 mm |
Wall Thickness | 1.1 mm |
Shell, Collector, and Guard Material | Shonka air-equivalent plastic C552 |
Venting | Vented to the ambient via the flexible vent tube surrounding the triaxial cable |
Stem | Non-removable 0.64 cm OD x 5.6 cm long, black anodized aluminum |
Nominal Collection Efficiency | 1.000 |
Maximum Polarizing Potential | 1000 V |
Nominal Inherent Leakage Currents | 10-15 A |
Low-Noise Triaxial Cable | 50 ohms, 29 pF/ft, 1.5 m long |
Signal Connector | Triaxial BNC plug (2-Lug, male pin); others available upon request |
High Voltage Connector | Integral with triaxial connector (shell of chamber is common with connector body) |
Waterproof | Yes |
Product Standards | CE0413, Designed to meet IEC 60601-1, IEC 60731 |
Model 14SL Schematic
Specifications subject to change without notice.